Objective:
To evaluate the accuracy and safety of pedicle screw placement using a new self-guided pedicle tap.
Methods:
According to the anatomical characteristics of the pedicle, a new self-guided pedicle tap was developed. Six adult spinal specimens including 4 males and 2 females were selected and tapped thread on the right and left sides of each pair of pedicles from the same segment T1 to L5 with traditional taps (control group) and new self-guided pedicle taps (experimental group), respectively. And the pedicle screws were placed. The screwing time was recorded and compared between two groups. CT scanning was completed to observe the accuracy and safety of the screw placement according to the Heary classification of imaging results.
Results:
The screwing time of thoracic and lumbar vertebrae in the experimental group were (5.87±1.25) min and(5.45±1.67) min, respectively. While those in the control group were (6.12±1.69) min and (6.22±2.13) min, respectively. Then there was no significant difference in screwing time of thoracic and lumbar vertebrae between two groups (P>0.05). The Heary grade of the pedicle screw showed that Heary gradeⅠand Heary gradeⅠ+Ⅱwere respectively 86 (84.31%) and 96 (94.12%) in the experimental group, 72 (71.29%) and 84 (83.17%) in control group, and the difference between two groups was statistically significant (P<0.05).
Conclusion:
The new self-guided pedicle tap can accurately and safely insert the thoracolumbar pedicle screw, with low cost, easy operation, and good clinical application value.
Keywords:
Anatomy; Feasibility studies; Pedicle screws; Spine.