Fluoroscopy time analysis of a prospective, multi-centre study comparing robotic- and fluoroscopic-guided placement of percutaneous pedicle screw instrumentation for short segment minimally invasive lumbar fusion surgery


Background:

As minimally invasive spine surgery becomes more widespread, concerns regarding radiation exposure to surgeons and patients alike have become a growing concern. Robotic guidance has been developed as a way to increase the accuracy of instrumentation while decreasing radiation burden.


Methods:

A retrospective analysis of a large, multi-centre, prospective study comparing robotic-guided (RG) to fluoroscopic-guided (FG) (Multi-centre, Partially Randomized, Controlled Trial of MIS Robotic vs. Freehand in Short Adult Degenerative Spinal Fusion Surgeries) was performed to evaluate for differences in radiation exposure between study groups.


Results:

RG was associated with 78.3% (p < 0.001) and 79.8% (p < 0.001) reduction in total and per screw fluoroscopy times, respectively, as compared to FG. RG was also associated with a 50.8% (p < 0.001) reduction in total operative fluoroscopy time.


Conclusions:

RG was associated with significantly lower fluoroscopy times compared to FG. This suggests that utilization of robotic navigation systems may result in decreased operative radiation exposure, which is a growing concern for surgeons performing minimally invasive spine surgery.


Keywords:

fluoroscopy; lumbar fusion; robotic spine surgery.