Study design:
Retrospective case series.
Objectives:
To investigate the influence of screw size on achieving bone fusion in posterior lumbar interbody fusion (PLIF).
Methods:
In total, 137 consecutive patients with L4 degenerative spondylolisthesis who underwent single-level PLIF at L4-L5 were evaluated. Factors investigated for their contribution to bone fusion included: 1) age, 2) sex, 3) body mass index, 4) bone mineral density, 5) intervertebral mobility, 6) screw diameter, 7) screw length, 8) screw fitness in the pedicle (%fill), 9) screw depth in the vertebra (%depth), 10) screw angle, 11) facetectomy, 12) crosslink connector, and 13) cage material.
Results:
Bone fusion was confirmed in 88.2% of patients. The comparison between fusion (+) and fusion (-) groups showed no significant differences in screw size. The %fill and %length were significantly greater in the fusion (+) group than in the fusion (-) group (%fill: 58.5% ± 7.5% vs 52.3% ± 7.3%, respectively, P = .005; %depth: 59.8% ± 9.7% vs 50.3% ± 13.8%, respectively, P = .025). Multivariate logistic regression analysis revealed that %fill (odds ratio [OR]= 1.11, P = .025) and %depth (OR = 1.09, P = .003) were significant independent factors affecting bone fusion. Receiver operating characteristic curve analyses identified a %fill of 60.0% and a %depth of 54.2% as optimal cutoff values for achieving bone fusion.
Conclusions:
Screw size should be determined based on the screw fitness in the pedicle (%fill > 60%) and screw insertion depth in the vertebral body (%depth > 54.2%) according to individual vertebral anatomy in L4-L5 PLIF.
Keywords:
bone fusion; lumbar degenerative spondylolisthesis; pedicle screw; posterior lumbar interbody fusion; screw insertion depth.