Background:
Posterior and lateral techniques have been described as approaches to sacroiliac joint arthrodesis. The purpose of this study was to compare the stabilizing effects of a novel posterior stabilization implant and technique to a previously published lateral approach in a cadaveric multidirectional bending model. We hypothesized that both approaches would have an equivalent stabilizing effect in flexion-extension and that the posterior approach would exhibit better performance in lateral bending and axial rotation. We further hypothesized that unilateral and bilateral posterior fixation would stabilize both the primary and secondary joints.
Methods:
Ranges of motion (RoMs) of six cadaveric sacroiliac joints were evaluated by an optical tracking system, in a multidirectional flexibility pure moment model, between ± 7.5 N-m applied moment in flexion-extension, lateral bending, and axial rotation under intact, unilateral fixation, and bilateral fixation conditions.
Results:
Intact RoMs were equivalent between both samples. For the posterior intra-articular technique, unilateral fixation reduced the RoMs of both primary and secondary joints in all loading planes (flexion-extension RoM by 45%, lateral bending RoM by 47%, and axial RoM by 33%), and bilateral fixation maintained this stabilizing effect in both joints (flexion-extension at 48%, lateral bending at 53%, and axial rotation at 42%). For the lateral trans-articular technique, only bilateral fixation reduced mean RoM of both primary and secondary sacroiliac joints, and only under flexion-extension loads (60%).
Conclusion:
During flexion-extension, the posterior approach is equivalent to the lateral approach, while producing superior stabilization during lateral bend and axial rotation.
Keywords:
Allograft; Arthrodesis; Bilateral; Compression; Distraction; Interference; Interposition; Range of motion; Unilateral.