Introduction:
Instrumentation with pedicle screws (PS) can compromise the adjacent neural structures. Triggered electromyography (tEMG) is a modality of intraoperative neuromonitoring, used to assist in the placement of these. The objective of the present study is to evaluate the reliability of this tool.
Method:
Retrospective review of patients underwent posterior lumbar fusions from January 2017 to December 2019, correlating postoperative CT images (postop CT) with tEMG results for each PS with a threshold of 10mA, establishing sensitivity and specificity of this tool. Diagnostic test and receiver operating characteristic curve were performed to evaluate the area under the curve.
Result:
A total of 275 PS were evaluated between L1 to S1; 5 PS showed concordance for an inadequate trajectory, while 10PS, with tEMG <10mA, were correctly positioned. The postoperative CT identified 17 pedicle gaps not noticed by the tEMG. Sensitivity and specificity were 23% (95% CI: 8-45) and 96% (95% CI: 93-98), respectively, with a PPV of 33.3% and a NPV of 93.6%. The area under the curve was 0.74 (95% CI: 0.62-0.86) with a cut-off point of 24mA, showing a sensitivity of 77% (95% CI: 0.55-0.92) and specificity of 69% (95% CI: 0.63-0.75).
Conclusion:
Given its low sensitivity, we do not recommend tEMG as the sole test in the verification of PS. We recommend using a cut-off point ≤8mA as it demonstrated the highest levels of sensitivity and specificity. We do not recommend using high thresholds to increase sensitivity.
Keywords:
Brecha pedicular; Electromiografía evocada; Especificidad; Evoked electromyography; Pedicle gap; Pedicle screw; Sensibilidad; Sensitivity; Specificity; Tornillo pedicular.